959D Leak Detector
2-4
DRAFT
9/13/02
MAXIMUM LEAK RATE Grouping
➀
FOCUS
Used for tuning the spectrometer tube by adjusting the focus
plate-to-chamber voltage.
➁
ION
A 10-position dial for adjusting the ion source chamber voltage.
Used for tuning the spectrometer tube.
1
3
4
5
6
7
8
9
10
2
➂
RESIDUAL
BACKGROUND
Kept in the RUN position except when
checking the background leak rate.
➃
COARSE ZERO
For large zero leak rate adjustments.
➄
FIL Grouping
Controls the filament in the ion source.
1 or 2 toggle switch Selects between ion source filaments
1 and 2.
ON/AUTO or OFF
toggle switch
ON/AUTO provides filament activation
when HI VAC OK is achieved. OFF dis-
ables filament emission.
EMIS
Adjusts the ion source emission voltage
when tuning the spectrometer tube.
➅
AUDIO
VOL outer ring: Controls the volume of
internal speaker, external speaker, or
headphones.
THOLD inner ring: Volume threshold for
an alarm.
➆
CAL
Adjusts amplifier gain to make the leak
rate reading agree with a calibrated leak.
➇
MAIN POWER
Main power control for the 959D.
➈
READY/TURBO
LED that indicates the Turbo or Macro-
Torr pump is at full operational speed.
➉
SENSITIVITY
Switches leak rate sensitivity between HI
and LO.
Used in conjunction with the RANGE
switch.
Figure 2-4 Secondary Controls and Indicators
Summary of Contents for 959D
Page 16: ...959D Leak Detector 1 2 D R A F T 9 1 3 0 2 Figure 1 1 Model 959D Leak Detector ...
Page 28: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 48: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 68: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 78: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 81: ......