959D Leak Detector
3-2
DRAFT
9/13/02
3.1.4 O-rings
Whenever installing an O-ring, follow these guidelines to prevent scratching of any sealing
surfaces:
❑
Wipe all O-rings clean with a lint-free cloth before installation to assure that no
foreign matter is present to impair the seal.
❑
Do not use grease or use other substances on O-rings that will come in contact with
the spectrometer tube.
❑
Do not use alcohol, methanol, or other solvents on O-rings.
To do so causes deterioration and reduces the O-ring’s ability to hold a vacuum.
❑
In allowable situations, apply a small amount of Apiezon
L grease and wipe the
O-rings “shiny” dry.
3.1.5 Cleaning Kit
The Varian Component and Spectrometer Tube Cleaning Kit, p/n 670029096, is
recommended for cleaning the spectrometer tube components, as well as for the fine
cleaning of other parts in the leak detector’s vacuum system such as valves and fittings.
Rinsing steps or high-temperature drying are not required following cleaning with the
VacuSolv used in this kit. Although appropriate precautions are advised, VacuSolv is com-
patible with most materials and does not contain toxic chemicals or CFCs.