959D Leak Detector
3-6
DRAFT
9/13/02
3.4.1 Removing the Ion Source
Replacement of the ion source should be accomplished when cleaning the spectrometer
tube or as soon as possible after a filament burns out. Tuning is required after replacement.
WARNING
When handling the ion source, wear powder-free butyl or
polycarbonate gloves. If you handle the ion source with bare
hands, wash them thoroughly and especially before smoking
or eating.
1. Shut down to the leak detector.
❑
Press the
VENT
button.
❑
Turn off the
MAIN POWER
switch on the front panel.
WARNING
Disconnect power from the 959D before performing any
maintenance procedure that requires physically disconnecting
any part of the system.
❑
Turn off the power switch on the AC power strip located at the bottom of the cart.
2. Remove the top cover as described in Section 3.3 “Removing the Cover” on page 3-5.
3. Locate and remove the connectors from the preamplifier and the ion source (Figure 3-3)
by grasping each plug and gently pulling straight off the contact pins.
If the connector cables are not labeled
ION SOURCE
and
PRE-AMP
, label them now.
Figure 3-3
Ion Source and Preamplifier Connectors
Preamplifier
Ion Source
Summary of Contents for 959D
Page 16: ...959D Leak Detector 1 2 D R A F T 9 1 3 0 2 Figure 1 1 Model 959D Leak Detector ...
Page 28: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 48: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 68: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 78: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 81: ......