959D Leak Detector
3-9
8. Remove the ion source flange cap by pressing down on the ion source center post
and
lifting the flange cap up and off the ion source as shown in the images of Figure 3-8.
Figure 3-8
Disengaging Flange Cap
9. If the flange cap and ion source come up together, hold one hand under the ion source
to catch it and then push down on the center post.
10. Remove the ion source from the spectrometer tube body as shown in Figure 3-9.
Figure 3-9
Removing the Ion Source