A.3 Terminology
The following terminology has application throughout this manual:
Flow
atm cc/sec
One cubic centimeter of gas per second at a pressure differential of
one standard atmosphere (760 torr at 0 °C).
atm cc/sec
One cubic centimeter of gas per second at ambient atmospheric
pressure and temperature (used interchangeably with
atm cc/sec
because the difference is insignificant for leak testing purposes).
Rate-of-Rise
In vacuum systems this is defined as the rate of increase of absolute
pressure per unit time, with the vacuum pump isolated from the
system, and is the sum of actual inleakage and internal outgassing.
Rate of rise is usually expressed in torr or microns (milliTorr) per
hour. The flow rate should be expressed in torr-liters/second.
Conversions
1 std cc/sec*
0.76 torr-liter/sec
1 torr-liter sec* 1.3 std cc/sec
1 std cc/sec
9.7 x 10
4
micron cubic feet per hour or
practically 10
5
micron CFH (
µ
CFH)
1
µ
CFH
practically 10
−
5
std cc/sec
*for practical purposes, equal
Numerical
Notation-Exponential
System
Most leak rates of commercial significance are very small fractions
of atm cc/sec. Therefore negative powers of ten are used as a
convenient system of numerical shorthand.
Table A-1 shows the relationship of exponents and multipliers (to the base 10) to the
arithmetic form, and the equivalent result.
Table A-1
Decimal Notation
Multiplier x 10
n
Arithmetic Form
Result
1 x 10
2
=
1 x 10 x 10
=
100
1 x 10
1
=
1 x 10
=
10
1 x 10
o
=
1
=
1
1 x 10
−
1
=
1 x 1/10
=
.1
1 x 10
−
2
=
1 x 1/10 x 1/10
=
.01
5 x 10
−
3
=
5 x 1/10 x 1/10 x 1/10
=
.005
5 x 10
−
7
=
5 x 1/10 x 1/10 x 1/10 x 1/10 x 1/10 x 1/10 x 1/10
=
.0000005
5 x 10
−
9
=
5 x 1/10 x 1/10 x 1/10 x 1/10 x 1/10 x 1/10 x 1/10 x 1/10 x 1/10
=
.000000005
Summary of Contents for 959D
Page 16: ...959D Leak Detector 1 2 D R A F T 9 1 3 0 2 Figure 1 1 Model 959D Leak Detector ...
Page 28: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 48: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 68: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 78: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 81: ......