GENERAL INFORMATION 

This equipment is destined for use by professionals. The user 
should  read  this  instruction  manual  and  any  other  additional 
information supplied by Varian before operating the equipment. 
Varian will not be held responsible for any events occurring due 
to  non-compliance,  even  partial,  with  these  instructions,  im-
proper  use  by  untrained  persons,  non-authorized  interference 
with  the  equipment  or  any  action  contrary  to  that  provided  for 
by specific national standards. 
The Turbo-V 700HT series pumps are turbo-molecular pumps 
for high and ultra-high vacuum applications and can pump any 
type of gas or gas compound. They are not suitable for pump-
ing  liquids  or  solid  particles.  The  pumping  action  is  obtained 
through  a  high  speed  turbine  (max.  42000  rpm)  driven  by  a 
high-performance  3-phase electric motor. The Turbo-V 700HT 
pumps  are  free  of  contaminating  agents  and,  therefore,  are 
suitable for applications requiring a "clean" vacuum. 
The following paragraphs contain all the information necessary 
to  guarantee the safety of the operator when using the equip-
ment. Detailed information is supplied in the appendix "Techni-
cal Information". 

This manual uses the following standard protocol:

 

 

!

!

!

!

 

WARNING!

 

The  warning  messages  are  for  attracting  the  attention  of  the 
operator  to  a  particular procedure or practice which, if not fol-
lowed correctly, could lead to serious injury. 

!

!

!

!

   CAUTION 

The caution messages are displayed before procedures which, 
if not followed, could cause damage to the equipment. 

NOTE 

The notes contain important information taken from the text. 

STORAGE 

When  transporting  and  storing  the  pumps,  the  following  envi-
ronmental requirements should not be exceeded: 

− 

temperature: from -20M to +70 MC 

− 

relative humidity: 0 - 95% (non-condensing) 

If  the  storage  time  is  long  start  the  pump  using  the  SOFT 
START mode (see control unit manual). 

PREPARATION FOR INSTALLATION 

The  pump  is  supplied  in  a  special  protective  packing.  If  this 
shows  signs  of  damage  which  may  have  occurred  during 
transport, contact your local sales office. 
When  unpacking  the  pump,  be  sure  not  to  drop  it  and  avoid 
any kind of sudden impact or shock vibration to it. 
Do  not  dispose  of  the  packing  materials  in  an  unauthorized 
manner.  The  material  is  100%  recyclable  and  complies  with 
EEC Directive 85/399. 

!

!

!

!

   CAUTION 

In  order  to  prevent  outgassing  problems,  do  not  use  bare 
hands to handle components which will be exposed to vacuum. 
Always use gloves or other appropriate protection. 

NOTE 

Normal exposure to the environment cannot damage the pump. 
Nevertheless, it is advisable to keep it closed until it is installed 
in the system, thus preventing any form of pollution by dust. 

 

INSTALLATION 

Do  not  install  or  use  the  pump  in  an  environment  exposed  to 
atmospheric  agents  (rain,  snow,  ice),  dust,  aggressive  gases, 
or in explosive environments or those with a high fire risk. Dur-
ing  operation,  the  following  environmental  conditions  must  be 
respected: 

− 

maximum pressure: 2 bar above atmospheric pressure 

− 

temperature: from +5 MC to +35 MC 

− 

relative humidity: 0 - 95% (non-condensing) 

In the presence of magnetic fields the pump must be protected 
using  a  ferromagnetic  shield. See the appendix "Technical In-
formation" for detailed information. 
The Turbo-V 700HT series pumps must only be used with one 
of  the  special  Varian  controllers  (series  969-9445,  969-9545, 
969-9872)  and  must  be  connected  to  a  primary  pump  (see 
"Technical Information"). 
 

!

!

!

!

   

CAUTION!

 

Do not remove the adhesive and protective cap before con-
necting the turbopump to the system. 
 
 

 

 

Summary of Contents for 969-9057

Page 1: ...UALE DI ISTRUZIONI BEDIENUNGSHANDBUCH NOTICE DE MODE D EMPLOI MANUAL DE INSTRUCCIONES MANUAL DE INSTRU ES BEDRIJFSHANDLEIDING INSTRUKSTIONSBOG BRUKSANVISNING INSTRUKSJON MANUAL OHJEK SIKIRJA ODHGIES C...

Page 2: ...Turbo V 700HT...

Page 3: ......

Page 4: ......

Page 5: ...RBOPUMP 25 TECHNICAL SPECIFICATION 26 INLET SCREEN INSTALLATION 28 HEATER BAND INSTALLATION 29 WATER COOLING KIT INSTALLATION 30 AIR COOLING KIT INSTALLATION 31 GAS PURGE VALVE INSTALLATION 32 VENT VA...

Page 6: ...seguenti condizioni ambientali temperatura da 20 QC a 70 QC umidit relativa 0 95 non condensante Se il tempo di immagazzinamento molto lungo avviare la pompa nel modo SOFT START vedere il manuale del...

Page 7: ...e aria o gas inerte esente da polvere o particelle La pressione di ingresso attra verso l apposita porta deve essere inferiore a 2 bar oltre la pressione atmosferica ATTENZIONE Evitare urti oscillazio...

Page 8: ...en k nnen ANMERKUNG Die Anmerkungen enthalten wichtige Informationen die aus dem Text hervorgehoben werden LAGERUNG Beim Transport und bei der Lagerung der Pumpen m ssen folgende klimatische Verh ltni...

Page 9: ...umpe noch eventuelle hei e Anschlu teile ber hrt werden denn es besteht Verbrennungsgefahr ACHTUNG F r die Bel ftung der Pumpe trockene staub und partikelfreie Luft oder Edelgase verwenden Der Eingang...

Page 10: ...ller respecter les conditions environnementales suivantes temp rature de 20UC 70UC humidit relative 0 95 non condensante Si le temps d emmagasinage est tr s long mettre la pompe en marche en mode SOFT...

Page 11: ...sonnes ATTENTION Pour le refoulement de l air de la pompe utiliser de l air ou du gaz inerte exempt de poussi re ou de particules La pression d entr e travers la porte pr vue cet effet doit tre inf ri...

Page 12: ...ones ambientales siguientes temperatura de 20TC a 70TC humedad relativa 0 95 no condensadora Si el periodo de almacenamiento es muy largo poner en marcha la bomba en modalidad SOFT START v ase el manu...

Page 13: ...culas La presi n de entrada a trav s de la puerta apropiada deber ser inferior a 2 bar por encima de la presi n atmosf rica ATENCION Ev tense golpes oscilaciones o bruscos desplazamientos de la turbob...

Page 14: ...bombas n o devem ser superadas as seguintes condi es ambientais temperatura de 20VC a 70VC humidade relativa 0 95 n o condensante Se o tempo de armazenagem for muito longo accionar a bomba em SOFT STA...

Page 15: ...acess rios durante as opera es de aquecimento A elevada temperatura pode causar les es s pessoas ATEN BO Para a sa da de ar da bomba utilizar ar ou g s inerte sem poeiras ou part culas A press o de e...

Page 16: ...atie die uit de tekst is gelicht OPSLAG Tijdens het transport en de opslag van de pompen moeten de volgende omgevingscondities aanwezig zijn temperatuur van 20RC tot 70RC relatieve vochtigheid 0 95 ni...

Page 17: ...ijk letsel veroorzaken ATTENTIE Gebruik voor de luchttoevoer naar de pomp lucht of inert gas zonder stof of vaste deeltjes De inlaatdruk via de hiervoor bestemde poort moet minder dan 2 bar boven de a...

Page 18: ...skade p udstyret BEM7RK Bem rkningerne indeholder vigtige yderligere oplysninger OPLAGRING Under transport og oplagring af pumpene skal f lgende omgivelsesbetingelser v re overholdt temperatur fra 20R...

Page 19: ...nvendes luft eller en inaktiv luftart der er fri for st v og partikler Indgangstrykket gennem den s rlige bning skal v re under 2 bar PAS P Undg sammenst d vibrationer eller bratte bev gelser i forbin...

Page 20: ...maskinskada skall uppst OBSERVERA Detta visar p viktig information i texten F RVARING F ljande krav p omgivningsf rh llanden g ller vid transport och f rvaring av pumpen temperatur fr n 20QC till 70Q...

Page 21: ...pumpen eller n gra tillbeh r under uppv rmningen f r att undvika br nnskador VIKTIGT Anv nd luft eller delgas fri fr n damm och partiklar f r luftning av pumpen Trycket vid inloppet f r vara h gst 2 b...

Page 22: ...t skades MERK Merknadene inneholder viktig informasjon som er hentet fra teksten LAGRING N r pumpene transporteres eller lagres m f lgende milj messige krav oppfylles temperatur fra 20MC til 70MC rela...

Page 23: ...til brannskader FORSIKTIG Bruk st v og partikkelfri luft eller inaktiv gass ved lufting av pumpen Trykket ved inngangen m ikke v re mindre enn 2 bar over atmosf risk trykk FORSIKTIG Unng st t svingnin...

Page 24: ...mi tekstiss k sitellyt t rke t tiedot VARASTOINTI Pumppujen kuljetuksen ja varastoinnin aikana t ytyy seuraavien ymp rist a koskevien ehtojen olla t ytettyin l mp tila 20PC 70PC suhteellinen kosteus 0...

Page 25: ...in kuumentamisen aikana Korkea l mp tila voi saada aikaan henkil iden loukkaantumisia HUOMIO P stett ess pumppuun ilmaa tulee k ytt ilmaa tai jalokaasua joissa ei ole p ly tai hiukkasia Sis njohtavan...

Page 26: ...keu SHMEIWSH SHMEIWSH SHMEIWSH SHMEIWSH Oi shmei seij peri coun shmantik j plhrofor ej pou coun apospaste ap to ke meno APOQHKEUSH APOQHKEUSH APOQHKEUSH APOQHKEUSH Kat th di rkeia thj metafor j kai th...

Page 27: ...a mpore na prokal sei proswpik j bl bej PROSOCH PROSOCH PROSOCH PROSOCH Gia na diocete sete me a ra thn antl a crhsimopoie ste a ra adran j a rio kaqar ap sk nh lla stoice a H p esh eis dou diam sou...

Page 28: ...ollowing envi ronmental requirements should not be exceeded temperature from 20M to 70 MC relative humidity 0 95 non condensing If the storage time is long start the pump using the SOFT START mode see...

Page 29: ...articles for venting the pump The pressure at the vent port must be less than 2 bar above atmospheric pressure CAUTION Avoid impacts oscillations or harsh movements of the pump when in operation The b...

Page 30: ...gs The Macrotorr stators are in the form of self positioning machined discs with pumping channels and an opening restricted by the corresponding ro tor discs These are made of aluminium alloy During n...

Page 31: ...60 Operating position any Cooling requirements Natural air convection Forced air or water optional Operating ambient temperature 5q C to 35q C Coolant water flow 200 l h 0 89 GPM temperature 10q C to...

Page 32: ...TION 27 87 900 878 01 C Graph of nitrogen pumping speed vs inlet pressure Graph of compression ratio vs foreline pressure Graph of nitrogen throughput vs inlet pressure using the recommended mechanica...

Page 33: ...e inlet screen however does reduce the pump ing speed by about 10 The inlet screen is fitted in the upper part of the pump as shown in the figure The screen can be mounted on each pump The inlet scree...

Page 34: ...tact with the pump wall to ob tain fast and efficient heating Switch on the heater while the turbopump is in op eration In the event of turbopump overheat the heater will be automatically cut out by t...

Page 35: ...Two 6 mm 1 4 internal diameter rubber or plastic hoses from the water supply must be fitted to the two nozzles NOTE These hoses must be held on the respective noz zles using hose clips to avoid that t...

Page 36: ...specification air flow 25 l s 51 CFM input voltage 120 Vac 50 60 Hz maximum power 17 W dimensions 119 x 119 mm 4 69 x 4 79 inches Position the fan on the pump ensuring that the holes in the plate line...

Page 37: ...ge port cover as shown in the follow ing figure and then screw the gas purge valve with a torque of 2 5 Nm as shown in the following figure VENT VALVE INSTALLATION The vent valve mod 969 9833 allows t...

Page 38: ...INSTALLATION The vent device mod 969 9831 allows to avoid undesired venting of the pump during a temporary power failure adjustable time up to 36 min and enables an automatic vent operation Unscrew an...

Page 39: ...LANGE To connect the Turbo pump to the ISO inlet flange remove the outer ring and position the centering ring as shown in the figure Then fix the two flanges with the clamps or claws as shown in the f...

Page 40: ...mechanical oil sealed pump it is advisable to in stall a suitable trap between the turbopump and the fore vacuum pump in order to prevent oil backstreaming Connection C ELECTRICAL The turbopump is con...

Page 41: ...pheric pressure in the system 4 Turn off the Turbopump 5 When the Turbo V pump and the backing pump are stopped and the system is at at mospheric pressure for a better bearing pro tection it is advisa...

Page 42: ...ime 969 9833 Vent device with adjustable delay time 969 9831 Forepump DS 402 with 1 ph 3 ph universal motor 949 9382 Dry forevaccum pump MD 60 120 V 949 9411 Dry forevacuum pump MD 60 220 V 949 9412 T...

Page 43: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Page 44: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Page 45: ...an India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 1...

Reviews: