5-2
Parameter
Range
Default
Description
Last Vial
1-51
01
This is the last vial or blank number to be
run. If BLANK sample type is selected,
enter the last number of blank to be run.
For example if you want two blank to be
run enter first vial=1 and last vial =2.
Sample Volume
0-25 ml
05
If the sample Type selected is Water or
Blank, this is the volume removed from
the sample vial or Blank/Wash bottle and
transferred to the sparge vessel. Note: If
the sample type selected is Soil, a
minimum of 3 ml of dilution water must be
selected in order for Standards to be
accurately added. The volume may be set
in 1 ml increments
Dilution
Factor
0,2,5,10,
20,50,100
NO
The dilution factor shows the ratio of the
total volume of water sample and blank
water blended together to the volume of
24.3 mls., DF-NO is no dilution, DF-2 is 1
volume sample blended with 1 volume
blank water (50% dilution), DF-50 is 1
volume sample blended with 49 volumes
blank water (98% dilution). The “4
” key
is used to change the dilution factor.
Rinse Volume
0-25 ml
05
Volume of rinse water used to flush the
PAT sparge tube during the bake mode.
The volume may be set in 1 ml
increments.
#Rinses
0-20
01
Number of times the sparge tube will be
flushed, using the rinse volume above,
during bake mode.
Standard 1
YES/NO
NO
This parameter will enable or disable the
addition of 1 µl of standard #1 to the
Sample or Blank or into the Soil vial.
Standard 2
YES/NO
NO
This parameter will enable or disable the
addition of 1 µl of standard #2 to the
Sample or Blank or into the Soil vial.
S. PreHeat
Stir
YES/NO
NO
This parameter turns the stirring motor on
or off during the Soil PreHeat cycle. It can
be used to break up the solid sample.
Note: This is a soil only parameter.