6-6
Remove the sensor device from the vial gripper. (Gently pull it down until it slides
out.) Press any key to park the vial gripper.
C A U T IO N
If you fail to insert the sensor device into the vial gripper, the Archon will still
attempt to run the program, however, it will not be able to detect the sensor bar
and the following message will appear:
Calibration Failed.
Press Enter to Retry
- to exit.
Press ENTER and then insert the sensor device and run the program again.
Once the “Auto Calibration” is complete place a 40 mL VOA in Position 22.
Scroll down using the #2 key to <vial 22 position) and calibrate the vertical height.
Press the 7 key for Train. Press “7” key for train again, compare first and second
calibration numbers. If the numbers stay within 5-6 digits complete “water probe”
and “equilibrium block” using the #7 key to train.
After completing calibrations, you should run the “calibration test” again. If unit
does not fail, you are not ready to began running samples.
Note:
Knockoff clearance (01100) and standard clear (00970) should only be
checked for default setting. DO NOT TRAIN.