10
03-914647-00:4
Space
Requirements
Allow sufficient bench space to permit installation of workstations,
integrators, autosamplers and other peripheral GC equipment. The table
below lists the physical dimensions and weight of the GC and the
peripheral instruments which may be installed near it. Allow six inches of
space at the sides and rear of the GC to permit free air circulation.
The power cord, gas inlets and the optional Ethernet® communications
port are located on the rear panel of the instrument. The power switch is to
the rear of the top cover. Signal output connectors for peripheral
instrumentation such as autosamplers and integrators are located under
the left top cover of the instrument.
Table 1 Physical Dimensions of the 3800 GC System Components
Height
Width
Depth
Weight
Instrument
in.
cm
in.
cm
in.
cm
lb.
kg
3800 GC
20
51
26
66
22
56
95
43
8200 AutoSampler
20
51
6
16
16
41
24
11
Tekmar 3000 Purge and Trap
19
48
9
23
18
46
37
16.8
Archon Purge and Trap AutoSampler
17
43
21.5
55
21
53
80
36
Tekmar ALS-2016
27
69
15
38
15
38
35
13.1
Genesis Headspace AutoSampler
22
56
28
72
18
46
110
50
Star Chromatography Workstation
(computer with monitor, approximate
values)
17
43
17
43
21
53
35
16
Environmental Requirements
Pollution degree: 2
Humidity:
5% to 95% RH
Temperature: 10
°
to 40
°
C operating, -20
°
to 65
°
C
non-operating
The 3800 Gas Chromatograph is intended for indoor
use and certified for operation up to an altitude of
2000m, however, the 3800 is designed for use at higher
altitudes. Varian GCs should be protected from
corrosive chemicals or gases, dust/ particulate
accumulation, and direct venting of air conditioners,
heaters, furnaces or fans.