30
03-914647-00:4
Three types of ferrules are common: graphite, Vespel® or polyimide, and
graphite/Vespel.
•
Do not use graphite with ECD or GC/MS systems.
•
Ferrule ID is based on the diameter of the capillary column. See the
following table.
Part Number
Ferrule ID
(mm)
Fitting Size
(inches)
Column ID
(microns)
Polyimide
(10/pk)
Graphite
(10/pk)
Polyimide/
Graphite (10/pk)
No hole 1/16
″
fitting — 28-694503-01 — 28-694590-01
0.4 mm 1/16
″
fitting 50 - 250 28-694586-01 28-694583-01 28-694580-01
0.5 mm 1/16
″
fitting 320 03-908361-01 28-694561-01 28-694581-01
0.8 mm 1/16
″
fitting 530 28-694552-01 28-694042-01 28-694582-01
Capillary column connection to the Micro-TCD requires a special series of
graphite/Vespel ferrules (see table below).
Column ID Ferrule ID
Ferrule
Part Number
0.1 mm 0.4 mm CP85889
0.25 mm 0.4 mm CP85889
0.32 mm 0.5 mm CP470100
0.53 mm 0.8 mm CP470101