Column Installation and Conditioning
CP-3800 GC Getting Started Manual
37
3
Move the column nut and ferrule to within 5 cm of the column end.
Find the column insertion depth for your detector in the diagram
below. Measure the correct distance from the end of the column
and mark this distance. Refer to diagram below for correct
distances. DO NOT let the column nut and ferrule fall past this
mark. The marking medium can contaminate the ferrule.
Column
Insertion
Depths for
Varian
Detectors
N O T T O S C A L E
F ID /T S D
9 .5 cm
E C D
1 0.5 c m
T C D
3.9 c m
C ap illary C o lum n N u t
P /N 03 -9 4 9551 -0 0
F e rrule
F u sed S ilica C o lum n
P F P D
9 .7 c m
C
o
lu
m
n
In
s
e
rt
io
n
D
e
p
th
For the Micro-TCD, the end of the column should be flush with the end of the
ferrule when installed on the detector fitting.