M
AINTENANCE
Flame Ionization Detector
CP-3800 GC Getting Started Manual
69
Figure 6 Flame Ionization Detector Assembly
Replace the
Ferrule
After a period of prolonged use, the flame-tip ferrule may deteriorate and
crack. As a consequence, a leak may develop around the base of the flame
tip assembly that causes unstable baseline noise and reduced sensitivity. In
this situation, or when the detector is to be operated at temperatures above
350°C, you must replace the ferrule. Be sure to select the proper ferrule for
the application. The Vespel/graphite ferrule has a maximum temperature limit
of 350°C.
1
Signal Cable
2
Signal Probe
3
Ignitor Cable
4
Ignitor Probe
5
Collector
6
Ceramic Insulator
7
FID Tower
8
Flame Tip
9
Aluminum Crunch Washer
10
Detector Base
11
Column Ferrule
12
Column Nut
1
2
5
6
3
4
7
8
9
11
10
12