CP-3800 GC Operator’s Manual
93
Packed
Mode
When operating in the packed column mode, adjust the carrier gas flow to
applicable packed column flow rates (20 to 50 mL/min).
1041 Operation
This section contains the specific operating information and procedures
required for optimum performance of the 1041 on column injector. All
installations for both the instrument and the injector must be completed
before continuing further. Note that the 1041 injector is equipped with
Electronic Flow Control as standard.
1041 Injector
Method
The 1041 injector method component on the 3800 keyboard contains one
screen for setting the required injector temperature and viewing the status
information. The 1041 is an isothermal injector and cannot be temperature
programmed.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set(°C): 250, Actual(°C): 250
MIDDLE 1041
1041 Oven Power
ON
1041 Temperature (°C):
250