Method 1
Method 1
0.00
20.00
Col Press (psi): 10 Column Flow (mL/min): 15
Linear Velocity (cm/sec): 10 Total Flow (mL/min): 30
FRONT EFC TYPE 4, Page 2 of 2
Time (min)
Total Flow (mL/min)
Initial
30
0.01
500
1.50
30
Flow Rates
for
Operation
Recommended total flow rates differ, depending on the operating mode
used. Normally the total flow rate is set about twice the required column flow
rate. For example if the desired column flow rate is 15 mL/min, set the total
flow rate to 30 mL/min. When a valve switch is programmed, the total flow
should be increased to offset any effects of the pressure disturbance. In this
case program the total flow to 500 mL/min for one minute and then revert
back to the original total flow. The total flow program can be further
optimized by adjusting the duration of the higher flow rate or reducing the
total flow.