The third and fourth pages, located by pressing the adjustments softkey,
include detector-specific parameters not routinely adjusted. For the FID,
these pages allow you to select between a fast or slow detector time
constant (page 1) and to set detector gas flows if EFC is installed (page 2).
The normal setting for the time constant is fast.
Page 1 of the FID method:
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 5.15
FRONT FID Page 1 of 2
Oven Power:
ON
Temperature (°C):
300
Electronics:
ON
Range:
12
IGNITE
AUTOZERO
ADJUSTMENTS
The IGNITE softkey applies power to the FID ignitor filament for 5 seconds.
The AUTOZERO softkey applies an immediate zeroing of the detector signal.
Page 2 allows time programming events such as range and Autozero.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 5.15
FRONT FID Page 2 of 2.
Time
Range
Autozero
Initial
12
NO
5.00
12
YES
6.50
11
NO
IGNITE
AUTOZERO
ADJUSTMENTS