CP-3800 GC Operator’s Manual
109
The following page is accessed by pressing the Adjustments softkey.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 5.15
FRONT FID ADJUSTMENTS, Page 1 of 2
Time Constant:
Fast
Clear AUTOZERO
The second adjustments page is only accessible if detector EFC is installed.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Make-up Flow (ml/min): 25 H
2
Flow (ml/min): 30
Air Flow (ml/min): 300
FRONT FID ADJUSTMENTS, Page 2 of 2
Make-up Flow (ml/min)
25
H
2
Flow (ml/min)
30
Air Flow (ml/min)
300
Installation/Disassembly
At times, it is necessary to remove the detector components to inspect, to
clean, to replace parts, or to install another detector. Follow the
disassemble/reassemble instructions below for both the FID and the detector
oven.
Figure 13 is an exploded view of the FID. Refer to these figures as you
disassemble the detector.