110
03-914648-00:7
Figure 13 FID Exploded View
Disassemble
the FID
To disassemble the FID, proceed as follows:
1
Set the FID to 50 °C and wait for it to cool before disassembly. Turn the
detector electronics and detector oven OFF in the active method. Remove
the GC top covers.
2
Turn off the main GC gas supplies to the detector at the pneumatics panel.
These are the make-up, hydrogen and air supplies.
3
Disconnect the signal and ignitor cables from their probes on the detector
tower.
1 Signal
Cable
2 Signal
Probe
3 Ignitor
Cable
4 Ignitor
Probe
5 Collector
6 Ceramic
Insulator
7 FID
Tower
8 Flame
Tip
9
Aluminum Crunch Washer
10 Detector
Base
11 Column
Ferrule
12 Column
Nut
1
2
5
6
3
4
7
8
9
11
10
12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com