116
03-914648-00:7
6
Page forward to page 2 and turn the initial Autozero OFF. Set the bead
current to 2.400 A. Turn on the power to the bead by setting Electronics to
ON. Allow the bead to condition for at least 15 minutes.
7
Increase the bead current in steps of 0.100 Ampere, waiting at least 15
minutes for the bead to equilibrate between increases. When increasing the
bead current causes the TSD baseline value to go above 64 mV, as shown
at the top of the display, the bead has reached its ignition point. Allow the
bead current to remain at this setting for a minimum of 12 hours (preferably
overnight). This completes the initial conditioning of the TSD bead probe.
Now optimization of the bead probe for N/C selectivity is required.
Page 1 of the TSD method:
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 15.35
MIDDLE TSD Page 1 of 2
Oven Power:
ON
Temperature (°C):
300
Electronics
ON
Range:
12
Bead Current:
2.4
AUTOZERO
ADJUSTMENTS
The AUTOZERO softkey applies an immediate zeroing of the detector signal.