CP-3800 GC Operator’s Manual
117
Page 2 allows time programming events such as range changes.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 15.35
MIDDLE TSD Page 2 of 2
Time
Range
Autozero
Bead Power
Initial
12
NO
ON
5.00
12
YES
ON
6.50
11
NO
ON
AUTOZERO
ADJUSTMENTS
The following pages are accessed by pressing the Adjustments softkey.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 15.35
MIDDLE TSD ADJUSTMENTS, Page 1 of 2
Time Constant:
Fast
Clear AUTOZERO
The Clear Autozero softkey removes the Autozero offset so that you can
observe the detector background signal.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com