131
Page 1 of the TCD method:
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 120 Actual (°C): 120 Signal (mV): 10.835
Filament Current (mA): 220 Balance (%): -42
FRONT TCD, Page 1 of 2
TCD Oven Power:
ON
Temperature (°C):
120
Electronics
ON
Range:
0.5
Filament Temperature: 170
AUTOZERO
ADJUSTMENTS
The AUTOZERO softkey applies an immediate zeroing of the detector signal.
Page 2 allows time programming events such as range changes.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 120 Actual (°C): 120 Signal (mV): 10.835
Filament Current (mA): 220 Balance (%): -42
MIDDLE TCD, Page 2 of 2.
Time
Range
Autozero
Polarity
Initial
0.5
NO
Positive
5.00
0.5
YES
Positive
6.50
0.5
NO
Positive
AUTOZERO
ADJUSTMENTS
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com