CP-3800 GC Operator’s Manual
135
Reverse
Polarity
If you inject a standard or sample into the reference side of the TCD, peaks
appear in the negative direction. If you want to integrate or process these
peaks with a data system, you must convert the polarity of the peaks from
negative to positive. Reference the TCD section to set initial detector
polarity either Positive or Negative.
You can also time program when you would like the polarity to be
automatically reversed during a chromatographic run. Set a time-
programmed polarity reversal in the detector method section.
Adjust TCD Carrier Gas Flow Rates
All 3800 GCs equipped with TCDs are factory tested at the flow rate listed
below. Adjustments to the carrier gas flow rate are made at the pneumatics
panel of the GC.
Carrier Gas:
Helium
Inlet Pressure:
80 psig
Purity:
99.999%
Column Flow Rate:
30 mL/min
Reference Column Flow Rate: 30 mL/min
Note:
Tightly cap gas filters when not used. Prolonged exposure of the
contents of the filter to room air degrades the performance of the filters.