138
03-914648-00:7
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 2.15
FRONT ECD, Page 1 of 2
ECD Oven Power:
ON
Temperature (°C):
300
Electronics:
ON
Range:
10
AUTOZERO
ADJUSTMENTS
The AUTOZERO softkey applies an immediate zeroing of the detector signal.
Page 2 allows time programming events such as range and Autozero.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 2.15
FRONT ECD, Page 2 of 2.
Time
Range
Autozero
Initial
10
YES
5.00
10
YES
6.50
1
NO
AUTOZERO
ADJUSTMENTS
ECD
Adjustments
There are ECD adjustments which should be set before operating the
detector. These are located by pressing the Adjustments softkey in the ECD
method.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com