146
03-914648-00:7
6
Position the tower assembly on the detector base and install the two 8-32 x
3-inch tower mounting screws loosely in the tower assembly.
Note:
Do not cross-thread the tower mounting screws.
7
Insert the installation tool into the detector tower assembly (located over the
detector cell). Engage the hex nut at the lower end of the cell assembly.
Tighten the cell assembly into the base to provide a leak-free seal.
Do not
over-tighten.
8
Alternating between the two 8-32 x 3-inch tower mounting screws, evenly
tighten the tower into place.
9
Carefully install the signal and pulser probes in the tower arms. The signal
probe (bottom) has a twist-lock cable connector and a narrow clip to engage
the collector cylinder. The pulser probe (top) has a threaded cable connector
and a wide clip to engage the foil cylinder. Insert the probes into the tower
arms with the keys on the probes lined up with the notches in the tower
arms. If you encounter any resistance during the insertion of the probes,
check that the probes are being installed in the correct positions.
10
Install the insulated detector tower cap. The label shown in
Figure 21 is factory installed on a detector tower registered as a General
License Device.
11
Connect the pulser and signal cables to the appropriate probes of the ECD.
Refer to the designation on each cable connector.
12
Cable connections for the operation of the ECD require installation of the
ECD Electrometer Board.
13
Connect the gas supplies at the rear of the GC.
14
Replace the GC top cover and the detector cover.
15
Install the “CAUTION RADIOACTIVE MATERIAL” label (P/N 31-000347-00)
in a location on the front of the GC that is clearly visible. This label describes
the ECD radioactive isotope and the amount of radioactivity (Figure 22).
16
Return the installation tool to the ECD case. You will use this tool each time
you remove the ECD. Retain the ECD case should you need to store the
ECD or to return the detector to Varian. See the Radiation Safety Manual.