CP-3800 GC Operator’s Manual
151
TCD Test Mix
Column:
Non-polar, e.g., 1µ film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier:
Helium, 5 mL/minute
Injection:
1079, 200
°
C, 1 µL splitless
Temperature Program:
80
°
, hold 1 minute, 20
°
/minute to 200
°
, hold 1 minute
Detector:
220
°
C, Range 0.05, Filament Temperature 270
°
±
20
°
, Filament current 200
±
10 mA