152
03-914648-00:7
ECD Test Mix
Column:
Non-polar, e.g., 1µ film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier:
Nitrogen, 5 mL/minute
Injection:
1079, 200
°
C, 1 µL splitless
Temperature Program:
80
°
, hold 1 minute, 20
°
/minute to 200
°
, hold 1 minute
Detector:
300
°
C, Range 1, Capillary current setting
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com