154
03-914648-00:7
PFPD S Mode
Column:
Non-polar, e.g., 1µ
film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier:
Nitrogen, 2 mL/minute
Injection:
1079, 220
°
C, 1 µL splitless of a 1:30 dilution of FPD test mix in iso-octane
Tempe
r
ature Program:
80
°
, hold 1 minute, 20
°
/minute to 200
°
, hold 1 minute
Detector:
300
°
C, Range 10, Air1 17 mL/min., Hydrogen 14 mL/min., Air2 10 mL/min.
Gate Delay 6 msec., Gate Width 20 msec, Trigger level 200 mV
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com