8
03-914648-00:7
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
EDIT HEATED ZONES
Zone
Type
Temp Limit
Location
Coolant
Col Oven
Col Oven
450 (°C)
Col Oven
None
1
1079
450 (°C)
Front
LCO
2
2
1079
450 (°C)
Middle
LCO
2
3
None
450 (°C)
Rear
None
4
FID
450 (°C)
Front
None
5
FID
450 (°C)
Middle
None
6
None
450 (°C)
Rear
None
Save and Exit
The following are the available choices of heated zone type, temperature
limit, location, and coolant.
Default
Range
Heat Zone
Col oven 1,2,3,4,5,6
Type None
1177,1079,1041,1061,1093
TCD, FID, TSD, ECD, PPFD
Large Valve Oven, Small Valve
Oven, Methanizer, SPT
Temp Limit ( °C)
450 °C
50 - 450 °C
1
Location
Front
Front, Middle, Rear
Coolant None None,
LN
2
, LCO
2
EFC
The 3800 GC can be configured with either manual pneumatics or electronic
flow control (EFC). Nine types of EFC modules exist, and each type is
associated with a specific GC inlet system or detector. The Setup menu
identifies each EFC type and its associated inlet system or detector by
location (front, middle, rear). Below is a description of each EFC type,
display, and associate
1
d parameters. When using AutoCalibrate for injector
EFC, verify that the inlet pressure to the EFC module is at least 60 psi, a
column and septa must be installed for this test.
1
Large and small valve ovens are limited to 350
°
C. Micro-TCD is limited to 110
°
C.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com