166
03-914648-00:7
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
AUTOMATION MODE SELECT
Automation Mode: Manual Injection
Note: Selecting Single Method or Method
Sequence Automation will disable remote control.
The automation mode is selected using the increment or decrement key on
the 3800 keyboard. Press the Enter key to activate the required automation
mode. If Single Method or Method Sequence automation is chosen, the
corresponding LED will light in the automation section of the 3800 keyboard.
An automation method may be suspended by pressing the
Suspend
key in
the automation section of the 3800 keyboard. To resume the automation
method press the
Resume
key.
If the 3800 Stop key is pressed while automation is in progress the
automation method will be terminated.
Automation Methods
Local 3800 automation methods differ depending on whether or not an
addressable device is detected. Addressable devices are connected to one
of the serial interface ports on the top of the 3800. These ports are labeled
as SID1 and SID2 (Sample Introduction Device 1 and 2). When the 3800 is
powered up it monitors these ports to determine if an addressable device is
present. If a device such as an 8400 AutoSampler is detected, the 3800
defaults to the addressable automation mode. Note that the default port for
an 8400 AutoSampler is SID1. If there are no devices connected to SID1 or
SID2, the 3800 defaults to the non-addressable automation mode. The
primary difference between addressable and non-addressable automation is
that the addressable methods include a sample range. For example, the
8400 AutoSampler contains a sample range of 0 to 99 sample vials.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com