CP-3800 GC Operator’s Manual
171
current injection is complete. Automation will resume after the priority
sample run is completed.
8400 AutoSampler Control
If an 8400 AutoSampler is detected during the 3800 power up, it will be
available to program in the 3800 method. The 8400 is programmed in the
Sample Delivery section of the 3800 method.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Current Vial: --- Injection Mode: Standard S/SL
8400
Injection Mode: Standard S/SL
Sample Penetration Depth (%): 90
Solvent Penetration Depth (%): 90
Prep-Ahead Delay (min): 0.00
Default Clean
Clean Mode
The 8400 has six injection modes: Standard On-Column, Standard
Split/Splitless, Volatile, Neat, Viscous and User Defined. These modes are
explained in detail in the 8400 Operator’s manual. The sampling mode is
selected using the increment or decrement keys on the 3800 keyboard.
•
The
standard On-Column and Split/Splitless
modes are the default and
typically yields good results in most applications. This mode uses a 1.0
µ
L air plug after the sample.
•
The
volatile
sampling mode is used for sampling solvents with high vapor
pressures at room temperature.
•
The
neat
sampling mode is used for pure samples without dilution.
•
The
viscous
sampling mode is used for samples with a higher viscosity
than water.
•
User
defined mode allows the user to specify additional injection
parameters.