The 3800 has a status screen which is dedicated to automation status. This
screen is accessed by pressing the status key on the 3800 keyboard and
using the Page Down key to access Local Automation Status. The
information displayed on the screen varies depending on whether
addressable or non-addressable automation is active.
Addressable Automation status screen (8400 Automation)
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Automation Status
Automation State: Ready
Injection Number: 1
Table Line Number: 1
Sample Vial: 1
Plunger Stroke Count 126
Prep-Ahead Delay (min) 0.00
The Automation Status field indicates whether the automation status is
ready, running or suspended. Injection number refers to the specific 8400
vial that is being sampled. Table line number indicates the line number if a
method sequence is being run and sample vial refers to the specific 8400 vial
being sampled (0 - 99). Note that table line number is blank if Single Method
automation is active.
Running Priority Sample
The priority sample is a powerful tool for temporarily interrupting the
execution of automation in order to run an urgent sample. While Local
Automation is running, you may select the Priority Sample Option. You will
be presented with options for which AutoSampler Vial Position and Method
you wish to use. After entering the pertinent information, pressing the Run
Now softkey will cause the priority sample to be run after the current injection
is complete. Regular automation will continue from where it was interrupted
after the Priority Sample is complete. If you select the Priority Sample on the
last run of either Single Method or Method Sequence automation, the system
will return to the beginning of Automation and continue. Press the Stop key
to abort running the same Automation a second time.