CP-3800 GC Operator’s Manual
11
Type 11
EFC
Type 11 EFC is a detector EFC module specifically designed for the Flame
Ionization Detector (FID). It is a flow controlled system with three
independent channels, channel 1 for make-up flow, channel 2 for hydrogen
flow and channel 3 for air flow. Note that the only parameter than can be
changed in the type 11 Setup screen is the choice of make-up gas. This
can be set to Helium (He), Nitrogen (N
2
) or Argon (Ar).
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
EDIT FRONT EFC (Type 11)
Detector:
FID
Channel 1 Gas Type:
He make-up
Channel 2 Gas Type:
H
2
Channel 3 Gas Type:
Air
AutoCalibrate
Save and Exit
For this Parameter …
Enter…
Channel 1 Gas Type
He, N
2
or Ar.
Type 11 EFC has the following softkeys:
Press this softkey
If you want to…
AutoCalibrate Automatically
calibrate the EFC system. When it is finished, a
screen will display either a successful completion or detected
problem message. No other GC functions can be performed
during AutoCalibration.
Note that AutoCalibration should be done
on a semiannual basis, or whenever an EFC module is changed.
Save and Exit
Save the changed parameters and reinitialize the 3800.