03-914648-00:7
Type 14
EFC
Type 14 EFC is a detector EFC module specifically designed for the
Thermal Conductivity Detector (TCD). It is a flow controlled system with two
independent channels, channel 1 for make-up flow and channel 2 for
reference flow (or make-up flow if two capillary columns are used with the
TCD). The gas type for both channels can be set to Helium (He), Nitrogen
(N
2
) or Argon (Ar).
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
EDIT FRONT EFC (Type 14)
Detector:
TCD
Channel 1 Gas Type:
He make-up
Channel 2 Gas Type:
He reference
AutoCalibrate
Save and Exit
For this Parameter …
Enter…
Channel 1 Gas Type
He, N
2
or Ar. Make-up.
Channel 2 Gas Type
He, N
2
or Ar. Make-up or reference.
Type 14 EFC has the following softkeys:
Press this softkey
If you want to…
AutoCalibrate Automatically
calibrate the EFC system. When it is finished, a
screen will display either a successful completion or detected
problem message. No other GC functions can be performed
during AutoCalibration.
Note that AutoCalibration should be done
on a semiannual basis, or whenever an EFC module is changed.
Save and Exit
Save the changed parameters and reinitialize the 3800.