CP-3800 GC Operator’s Manual
15
Type 15
EFC
Type 15 EFC is a detector EFC module specifically designed for the Pulsed
Flame Photometric Detector (PFPD). It is a flow controlled system with
three independent channels; channel 1 for air 1 flow, channel 2 for
hydrogen flow and channel 3 for air 2 flow. There are no user selectable
parameters with EFC type 15.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
EDIT FRONT EFC (Type 15)
Detector:
PFPD
Channel 1 Gas Type:
Air 1
Channel 2 Gas Type:
H
2
Channel 3 Gas Type:
Air 2
AutoCalibrate
Save and Exit
Type 15 EFC has the following softkeys:
Press this softkey
If you want to…
AutoCalibrate Automatically
calibrate the EFC system. When it is finished, a
screen will display either a successful completion or detected
problem message. No other GC functions can be performed
during AutoCalibration.
Note that AutoCalibration should be done
on a semiannual basis, or whenever an EFC module is changed.
Save and Exit
Save the changed parameters and reinitialize the 3800.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com