16
03-914648-00:7
Type 16
EFC
Type 16 EFC is a detector EFC module specifically designed for the
Thermal Conductivity Detector (TCD) when used with Hydrogen gas. It is a
flow controlled system with two independent channels, channel 1 for make-
up flow and channel 2 for reference flow (or make-up flow if two capillary
columns are used with the TCD). The gas type for either channel can be set
to make-up or reference.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
EDIT FRONT EFC (Type 16)
Detector:
TCD
Channel 1 Gas Type:
H
2
make-up
Channel 2 Gas Type:
H
2
reference
AutoCalibrate
Save and Exit
For this Parameter …
Enter…
Channel 1 Gas Type
Make-up or reference
Channel 2 Gas Type
Make-up or reference
Type 16 EFC has the following softkeys:
Press this softkey
If you want to…
AutoCalibrate Automatically
calibrate the EFC system. When it is finished, a
screen will display either a successful completion or detected
problem message. No other GC functions can be performed
during AutoCalibration.
Note that AutoCalibration should be done
on a semiannual basis, or whenever an EFC module is changed.
Save and Exit
Save the changed parameters and reinitialize the 3800.