18
03-914648-00:7
Sample
Delivery
Setup
This section has parameters for an autosampler, if installed, and valving.
Autosampler
In this section you configure which injectors you want the 8400 to injector into. You
also select the syringe capacity.
Valves
This section configures the seven external events available on the 3800 that are
typically used to control the CP-1177 and 1079 split valve with manual pneumatics
(non-EFC) or actuators for sampling or switching valves. Each valve is identified in
SETUP by a name that indicates its application. Names are selected using the INCR
and DECR keys from a table of 22 names. When a name is chosen this also defines
the default states of the valve. These states will be visible when the valve is being
programmed in the valve table section of the 3800 method.
Below is an example of a 3800 equipped with a manual pneumatics 1079
injector, a CP-1177 injector and two automated gas sampling valves (GSV).
Valve numbers 1 and 2 control the GSVs, and 3 and 4 control the CP-1177
and 1079 split valves. Each split valve type is identified based on the
location of its associated injector.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
EDIT VALVES, Page 1 of 2
Number
Valve Type
1
Gas Sampling Valve
2
Gas Sampling Valve
3
Front Split Valve
4
Rear Split Valve
Save and Exit