24
03-914648-00:7
EDIT Method
Key
The EDIT key allows you to select one of the eight methods to edit. The
active method is the default selection to edit, which allows quick access to
modifying the active method. The components available to edit are those
that have been configured earlier in the Instrument SETUP section. Below is
an example of the edit screen showing Method 1 as the active method. If
you want to edit another method, use the INCR/DECR keys to toggle to
another method, then press ENTER. The Injector Method screen will appear
first. If there is more than one Injector installed, a menu page will appear
instead, so that the desired injector may be selected for editing. Method
components appear in the following order: Injector, Flow/Pressure, Column
Oven, Detector, and Output. If you want to go directly to one of the method
sections, press the appropriate GC Control key. To access the Valve Table
or local 8200 AutoSampler method, press the Sample Delivery key.
Note:
The method parameters that appear on the display represent the current
EDIT method and not necessarily the ACTIVE method. The edited method
parameters will only be implemented when activated. Always remember to
reactivate the active method if you have made changes to it.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
EDIT METHOD
Edit Method:
Method 1