34
03-914648-00:7
1079 Injector
The 1079 has three pages:
Page Description
1
Temperature control and programming
2
Oven power and cryogenic parameters (if configured in Setup)
3
Split state programming (if type 1 EFC is installed)
If the 1079 has manual pneumatics, then the split valve is
programmed in the valve table located in SAMPLE DELIVERY.
The following are typical method parameters for a 1079 with cryogenics and
temperature programming.
Temperature programs are entered in page 1.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 50 Actual (°C): 50 Split Ratio: OFF
FRONT 1079, Page 1 of 3
Step
Temp (°C)
Rate (°C/min) Hold (min)
Total (min)
Initial
50
-
0.00
0.00
1
200
150
15.00
16.00
Default
Range
Step
Initial
[1 - 4]
Initial temperature (
°
C):
50
[-99 - 450
°
C]
Rate (
°
C/min):
X
[1 - 200]
Hold Time
X
[0.01 - 999.99]
Total X
[0.01 - 999.99]