36
03-914648-00:7
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 50 Actual (°C): 50 Split Ratio: OFF
FRONT 1079, Page 3 of 3.
Time
Split State
Split Ratio
Initial
OFF
OFF
0.75
ON
100
Default
Range
Time (min)
Initial
[0.00 - 999.99]
Split State
Off
[On/Off]
Split Ratio
100
[Off, 1 to 10,000]
If the injector is equipped with manual pneumatics, then the injector split
state is programmed in the VALVE TABLE of the SAMPLE DELIVERY
section.
Isothermal
(1041 / 1061)
Injector
The 1041 and 1061 isothermal injectors display one screen to set the
temperatures or view the status.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 250 Actual (°C): 250
MIDDLE 1041
1041 Oven Power:
ON
1041 Temperature (°C)
250
Default
Range
1041 Oven Power:
Off
[On/Off]
1041 Temperature (
°
C):
50
[50 - 450
°
C]