CP-3800 GC Operator’s Manual
43
A second page allows establishing optional column coolant parameters and
the stabilization time.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 50 Actual (°C): 50
Stabilization Time (min): 2.00 Column Oven: On
COLUMN OVEN, Page 2 of 2
Stabilization Time (min)
2.00
Coolant On/Off
On
Enable Coolant at (°C)
50
Coolant Time-out (min)
20.00
Default
Range
Coolant ON/OFF
Off
[On/Off]
Activate Coolant at (
°
C)
50
[30 - 450]
Coolant Time-out (min):
20.00
[0.00 - 999]
Stabilization Time (min)
0.50
[0.00 - 10.00 min]
DETECTOR
Key
The DETECTOR section allows you to modify the detector method
parameters and view the status of individual detectors. If more than one
detector is installed, then pressing DETECTOR will prompt you to select
from a menu of the installed detectors. The menu screen indicates the name
of the detector and its location.
Below is an example for a Flame Ionization Detector. Further details for
programming other detectors can be found in the detector section of this
manual. Each detector method consists of three or four pages. Page one
includes initial parameters for the detector oven temperature, turning on the
detector oven power and electronics, and selecting the range. Page two
establishes time programs to either change the range or Autozero the
detector at specific times. The adjustments page or pages are accessed by
pressing the adjustments softkey. For the FID, the first page of adjustments
allows you to select between a fast or slow detector time constant. The
normal setting for this parameter is fast.
If detector EFC is installed, then page 2 of adjustments allows setting the
appropriate flow rates for an FID.