44
03-914648-00:7
FID (Flame
Ionization
Detector)
The screen below indicates the set and actual detector oven temperature
and the FID analog signal level.
The method parameters, default settings, and range are:
Default
Range
Oven Power
Off
[On/Off]
Temperature (
°
C)
50
[50 - 450
°
C]
Electronics Off
[On/Off]
Range
12
[9, 10, 11, 12]
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 300 Actual (°C): 300 Signal (mV): 5.15
FRONT FID, Page 1 of 2
Oven Power:
ON
Temperature (°C):
300
Electronics:
ON
Range:
12
IGNITE
AUTOZERO
ADJUSTMENTS
The IGNITE softkey applies power to the FID ignitor filament for 5 seconds.
Note that there is a flame re-ignition diagnostic which re-lights the FID if a
flame-out is detected. The AUTOZERO softkey applies an immediate zeroing
of the detector signal. Note that this is a single Autozero event and not a
continuous Autozero.
Press the Page Down key to access the time programming events.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com