88
03-914648-00:7
The 1041 On-Column Injector
The 1041 on-column injector is designed for use with 0.53 mm ID capillary
columns or 1/8 - 1/4" packed columns. Switching from one mode to another
involves changing the injector and detector column mounting hardware. The
1041 is supplied with capillary mounting hardware as standard. If you wish
to operate the injector in the packed column mode then a packed column
adapter kit must be ordered (03-925588-91). This kit contains injector and
detector hardware to facilitate use of the injector with 1/8" metal packed
columns. In addition, adapter kits are available for 1/4" columns. Full
instructions on converting the 1041 to packed column use are included with
the packed column kit.
Automatic Start Switch
The automatic start switch is a spring loaded actuator that fits over, and is
aligned with the injection port of the 1041 universal injector nut. The GC run
is started when the actuator is depressed by the syringe barrel, or manually
pressed at the moment of sample injection. The GC run can also be
manually started by pressing START on the keyboard.
The injector nut and automatic start switch assembly may
be very hot during instrument operation and should not be
touched with unprotected hands.
Injector Assembly and Insert
A cross-sectional view of the 1041 Universal Injector with insert and 530µ ID
fused silica capillary column is shown in Figure 9.