CP-3800 GC Operator’s Manual
89
Figure 9 1041 Universal Injector with Insert
Column Installation
The following instructions apply to installing a 0.53 mm ID capillary column in
the 1041 injector. Note that the thin polymeric coating on fused silica
columns will give some protection against breakage; however, fused silica
columns are somewhat fragile and must be handled with care.
Mounting
the Column
During operation, the 0.53 mm ID capillary column hangs on the capillary
column holder in the column oven.
Note:
To prevent the column from unwinding, weave both of the column ends
through and around the column loop several times. This is particularly useful if
your column is not mounted in a cage.
1 Septum
Nut
2
Automatic Start Switch
3 Septum
4 Injector
Body
5 Injector
Insert
6 Capillary
Nut
1
2
3
4
5
6