DOC: ATP-03-925110-XX
DESC:
PAGE:
REV
14 of 16
TEST & CALIBRATION
2
PREPARED BY: HOWIE PLITSCH REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
08 - 01 - 00
Sax Makthepharack
Quick Test
- A Pre-Calibration test designed to weed out bad EFC modules. Other more advanced functions of this module are to "Test Controller" of individual
EFC’s. This is the "Controller" part of the screen and not covered in this section.
*
In the Group box, indicate the type of EFC’s in each group.
*
Set the Warm up Time to 1 minute.
*
Set the Flow Sensor Serial Number.
*
From the Quick Test Setup Pulldown menu, Select "Full Test". If IEFC’s are being tested, Set "Mux Test" button to on.
*
Click on the arrow in the upper left hand corner to start.
*
When finished, click the upper right hand arrow to return back to the main menu.