DOC: 03-925880-00
DESC:
PAGE:
REV
4 of 7
3800GC, Burn-In Test Procedure
2
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
07 - 30 - 99
F.Faili/M.Slaughter
1.7.5
For 1060, install Glass Insert 37-000813-00 in Burn-In. For 1061 install Glass Insert 03-918339-00 in Burn-In.
1.7.6
Remove & replace septum with aluminum seal.
1.7.7
Perform steps 1.7.1-1.7.6 for each 1041/60/61 injector.
1.8
DETECTORS
1.8.1
Press Detector key to get to menu.
1.8.2
Select Detector.
1.8.3
Turn On Detector oven.
1.8.4
Set temp. to 50
°
C.
1.8.5
Perform steps 1.8.1-1.8.4 for each detector.
1.9
FLOW/PRESSURE
1.9.1
Press Flow/Pressure key to get to menu (see diagram).
1.9.2
Set Pressure or Flow:
TYPE
PRESSURE
FLOW
Saturn
10 PSI
Manual
10 PSI
30-40mL/min
Type 1,5,21,25 EFC
10 PSI
Type 3,23 EFC
25mL/min
Type 4,24 EFC
25 PSI
1.9.3
For 1079 Injectors:
a.
Adjust Sport Valve Flow to 5-10 mL/min.
b.)
Verify flow coming from test column is 30-100 mL/min.
c.
Measure flow at EFC Vent tube, should be
≈
to Total Flow + Flow @ Sport Valve
±
10%.
d.
Press Activate key then Activate Now, (instrument must be in Ready State to Activate).
e.
If the above conditions are not met, determine cause and correct.
f.
Remove injector nut and Septum & install bake out nut and seal, T-8941.