Introduction and Initial Operation
Page 16
User Manual Micro-GC
Varian, Inc.
1HSAMPLE GAS
The Micro-GC is an analyzer built for the analysis of gases and vapors only. You are
advised to prepare a non-condensing gaseous samples standard sample for routine
check up of the instrument. Sample pressure should be between 0-100 kPa (0-15 psi),
the temperature between 0 and 40
C ± 5°C of the analyzer ambient temperature and it
must be filtered, preferably through a 5
m filter. Varian, Inc. ALWAYS recommends the
use of the external filter kit part number CP736729.
Click here for more
details about the heated sample line
.
Liquids will seriously damage the instrument and should be avoided!
Handling a sample
If possible, filter and dry the sample before introducing it to the Micro-GC. It is advised
to use an External Sample Filter Unit between the injector and the sampling device.
The Filter, male must be hand tightened into the Filter Female, followed by a 1/8th turn
with a 7/16" wrench. The arrow on the Filter Female should be directed towards the
Fingertight Fitting.
Replace the External Filter Unit
dat regular intervals.
Whenever possible remove moisture from samples introduced to the
Micro-GC.