Introduction and Initial Operation
Page 20
User Manual Micro-GC
Varian, Inc.
2HPOWER
The Micro-GC requires 12.0 VDC, 130W maximum.
Refer to the User manual power supply for more information part number CP501267.
Only use the power supply that has been supplied with the instrument.
SHUT DOWN PROCEDURE
When the instrument is shut down for more then a few days please,
carry out the procedure mentioned below
Create a method for all channels with the following settings:
Filaments switched OFF.
Column temperature set at 30°C.
Injector temperature set at 30°C.
Pressure set at 50 kPa (7 psi).
Wait until the temperature of the column and injector is < 40°C (to protect the column),
before switching off the Micro-GC.
Remove the carrier gas tubing and plug all the vents/carrier gas connections with 1/8”
brass nut or plastic cap.
When the instrument is going to be used again then follow the procedure