Introduction and Initial Operation
Varian, Inc.
User Manual Micro-GC
Page 21
LONG STORAGE RECOVERY PROCEDURE
Follow the recovery procedure below if your Micro-GC has been stored for a
long period of time
Remove the 1/8” brass nuts & plastic caps from all the vents/carrier gas
connections.
Connect carrier gas tubing and apply pressure on the Micro-GC, according
the pre-installation requirements.
Wait, at least 10 minutes, before switching ON the Micro-GC.
Check immediately if the detector filaments are switched OFF, switch OFF if
necessary .
Set the column(s) temperature to the maximum allowed temperature.
Condition the column module, preferably overnight. This will ensure that all
the water has been removed from the column module and no damage will
occur to the TCD filaments.