Reference
Varian, Inc.
User Manual Micro-GC
Page 49
SERIAL COMMUNICATION
To control the Micro-GC via serial communication, just connect the serial cable (supplied with
the instrument) to a free serial port of your computer configured as RS232, and plug the other
side into port COM 1 of the Micro-GC (see
Avoid cable lengths longer than 15 meter while this can cause communication
problems due to high communication speed (baud rate 115200 bits/sec).
The maximum number of Micro-GCs controlled by serial communication is two.
Choose Ethernet communication in case more than two Micro-GCs must be
controlled by one computer or setup a second computer with EZChrom Elite
installed. If RS485 is required, an RS232-RS485 converter (optional) must be used.
RS485 is only possible via two RS232-RS485 converters, one on Workstation
(computer) side and either one at Micro-GC side on Com 1.
Serial Cable
EZChrom Elite
490-GC
Com 1
490-GC
Com 1
Serial Cable
Com 1: RS232
Com 2: RS232
Summary of Contents for Micro-GC 490-GC