ProStar325 Service Manual
Publication date 8 May 2003
101
5.3
Removal and replacement procedures
Warning
Lamp surfaces and mounting brackets will become hot
during operation and will remain hot for some time after
being switched off. The temperatures are sufficient to cause
burns.
The UV lamps operate at high voltage. Contact with this
voltage can cause death, serious injury, or painful electric
shock.
Before changing either the UV or Visible lamps,
always
switch the instrument off, remove the power supply cable
and allow the lamps to cool.
Caution
Care must also be taken when removing lamps. Touching the
glass envelope of either the UV or Visible lamp will reduce its
efficiency. Never touch the glass surfaces of new lamps.
Aways handle a lamp by its base, or use a tissue or cloth.
Warning
Hazardous UV radiation is emitted by the UV lamp. This
radiation can cause serious damage to human eyes. Never
operate a UV lamp outside of its correct position in the lamp
turret
5.3.1
UV lamp source
Removal
•
Remove the front green door which will then expose a flat
panel.
•
Remove this flat panel by loosening the holding screw on
the top left hand corner. This will expose the UV lamp
housing which is located on the right hand side of the
opening.
•
Disconnect the lamp from the 3-way connector.
•
Locate the safety interlock latch and push it upwards as
far as it goes.
Entry
collimator
N2
StockCheck.com
Summary of Contents for ProStar 325 UV-Vis
Page 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Page 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Page 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Page 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...