ProStar325 Service Manual
Publication date 8 May 2003
63
3.21
Flowcell
The ProStar325 can be fitted with any one of four flowcells
(ordered separately). The customer’s chosen flowcell type is
packed internally in the detector. The four types are;
Part Number
Description
Volum
e
02 101818 00
FLOWCELL 9x0
ANALYTICAL
15ul
02 101819 00
FLOWCELL 9x1 PREP
15ul
02 101820 00
FLOWCELL 4x0.15 SUPER
PREP
15ul
02 101821 00
FLOWCELL 4x0 SEMI MICRO 1.6ul
All flowcells can be removed and re-installed as follows:
•
The flowcell should be removed with the connecting tubing
fitted. These must be removed outside the detector
compartment.
•
Access the flowcell compartment by removing the green
front door.
•
Remove the flat panel by loosening the single screw on the
top left hand corner.
•
Loosen the two thumbscrews which secure the flowcell.
•
Holding the flowcell by these two thumbscrews, carefully
remove it from its position.
Thumbscrew
StockCheck.com
Summary of Contents for ProStar 325 UV-Vis
Page 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Page 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Page 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Page 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...